Abstract :
[en] Background: “Less invasive surfactant therapy” (LIST), or tracheal instillation of surfactant through a small catheter in spontaneously breathing infants, is gaining popularity. Different catheters are currently used for this purpose: a nasogastric tube inserted with (LISA) or without (Take Care) Magill’s forceps, a 13 cm 16G adult angiocath (MIST), a 30 cm F4 angiography catheter (Stockholm). We developed a specific device by combining a F5 umbilical catheter and an intubation stylet (Liege). We aimed to compare those 5 devices using INSURE as a reference. Methods: 20 neonatologists from 4 institutions supporting different surfactant instillation policies intubated 2 manikin heads with the 5 catheters and an endotracheal tube in a predetermined random sequence. Water was flushed trough the catheter. Video review provided times between laryngoscope (T1) or catheter insertion (T2) in the mouth and water flowing from the trachea. Participants gave an ease of use score (range: 1-9) for each catheter. Results: Procedural times were longer with the Take Care method and shorter with the Liège device (Table). Failure rates were higher for LIST procedures than for INSURE. Take Care and LISA were rated as more difficult, while Liège, Stockholm and INSURE were considered easier. Conclusions: LIST procedures remain difficult, even on a manikin. The choice of catheter is important. A device combining the rigidity of a stylet with the soft distal end of an umbilical catheter is associated with procedures of shorter duration and is considered easier by neonatologists.
Disciplines :
Pediatrics
Author, co-author :
RIGO, Vincent  ;  Centre Hospitalier Universitaire de Liège - CHU > Service néonatologie (CHR)
Debauche, Christian;  Cliniques Universitaires Saint-Luc (Bruxelles) > Néonatologie
Maton, Pierre;  CHC-St Vincent, Rocourt > Néonatologie
BROUX, Isabelle ;  Centre Hospitalier Universitaire de Liège - CHU > Service néonatologie (CHR)
van Laere, David;  Universitair Ziekenhuis Antwerpen - UZA > Neonatologie
Language :
English
Title :
Devices for less invasive surfactant therapy: a manikin study
Publication date :
November 2016
Event name :
European Academy of Paediatric Societies Congress 2016
Event organizer :
ESPR, ESPNIC, EAP
Event place :
Genève, Switzerland
Event date :
21-25 octobre 2016
Audience :
International
Journal title :
European Journal of Pediatrics
ISSN :
0340-6199
eISSN :
1432-1076
Publisher :
Springer Verlag, Berlin, Germany
Volume :
175
Issue :
11
Pages :
1756
Peer reviewed :
Peer Reviewed verified by ORBi
Available on ORBi :
since 20 January 2017

Statistics


Number of views
171 (1 by ULiège)
Number of downloads
283 (0 by ULiège)

OpenCitations
 
10

Bibliography


Similar publications



Contact ORBi